Level-1: Final flux calculations
Level 1 is the final flux calculation with EddyPro, run with fluxrun, one run per setup period. Its FLUXNET output is the input of the post-processing chain. FP2026.1 contains it in full: 329 columns, listed in the section “L1 fluxes IRGA72” of the _INFO.md file.
The Level-1 fluxes are not quality-filtered, and FC carries no storage term. For analyses, use the columns of the post-processing chain (see QCF and Level 4.1).
Notebook: 99_IRGA72_L1_MERGED_2016-2026.09 merges the runs and checks the merged record.
The Level-1 fluxes of the open-path LI-7500 (IRGA75) are in progress for FP2026.2 and are not part of FP2026.1. They are merged, before the self-heating correction of Level 1.1, in 99_IRGA75_L1_MERGED_noSHC_2004-2017_2019.
The IRGA72 record
The merged record runs from 11 January 2016 15:45 to 30 September 2026 23:45 (TIMESTAMP_MIDDLE), 187,937 half-hours. In FP2026.1 the columns are empty before 11 January 2016.
IRGA72. A run covers one or more setup periods: 2017_1+2 is one run over setup periods 1 and 2 of 2017.
| Run | Period | Note |
|---|---|---|
2016_2 |
11 Jan - 26 May 2016 | start-up of the LI-7200 |
2016_3 |
26 May - 31 Dec 2016 | |
2017_1+2 |
2017 | |
2018_1 |
1 - 31 Jan 2018 | |
2018_2 |
31 Jan - 19 May 2018 | |
2018_3 |
19 May - 7 Jun 2018 | sonic only: H, no FC or LE (failed flow module) |
2018_4+2019_1 |
7 Jun 2018 - 10 Jan 2019 | |
2019_4 |
20 Mar - 31 Dec 2019 | second LI-7200 unit from here on |
2020 |
2020 | |
2021_1 |
1 Jan - 13 Dec 2021 | |
2022_2 |
7 Feb - 9 Mar 2022 | |
2022_3 |
9 Mar - 18 Nov 2022 | |
2022_4 |
18 Nov - 31 Dec 2022 | |
2023_1 |
1 Jan - 13 Mar 2023 | |
2023_2 |
13 Mar - 31 Dec 2023 | analyser inlet repositioned |
2024 |
2024 | |
2025 |
2025 | |
2026_1 |
1 Jan - 30 Sep 2026 |
Four setup periods have no IRGA72 run:
2019_2(11 to 17 January 2019): the LI-7200 was replaced on 11 January 2019 and both units were defective, so there are no usable gas fluxes. No sonic-only run forHhas been made.2019_3(17 January to 20 March 2019): the LI-7500 stood in for the LI-7200. Its fluxes belong to theIRGA75record (FP2026.2).2021_2and2022_1(13 December 2021 to 7 February 2022): the raw data cannot be converted (see Raw Data).
They form the two long gaps of the record: 10 January 15:15 to 20 March 18:45 2019 (69 days) and 13 December 2021 19:15 to 7 February 2022 10:45 (56 days).
FC, LE and H per year, before quality control. A full year has 17,520 half-hours (17,568 in a leap year).
| Year | FC |
LE |
H |
|---|---|---|---|
| 2016 | 16,378 | 16,558 | 16,876 |
| 2017 | 17,066 | 16,736 | 17,285 |
| 2018 | 15,899 | 15,846 | 17,194 |
| 2019 | 13,602 | 13,822 | 13,978 |
| 2020 | 16,125 | 16,232 | 17,433 |
| 2021 | 16,303 | 16,307 | 16,507 |
| 2022 | 15,225 | 15,226 | 15,572 |
| 2023 | 16,879 | 16,881 | 16,916 |
| 2024 | 16,943 | 16,944 | 16,993 |
| 2025 | 17,404 | 17,397 | 17,415 |
| 2026 (to Sep) | 13,039 | 13,039 | 13,082 |
Settings and inputs
The EddyPro settings of all runs, and how they differ between runs, are on Level-1: EddyPro settings.
- Time lags. The nominal lags and lag windows were chosen from the Level-0 runs; see the IRGA72 lag table on Level 0.
- Biomet input. Six meteorological variables (
TA,RH,PA,SW_IN,LW_IN,PPFD_IN) are passed to EddyPro, exported from the meteo products. The runs up to2025read the 2004-2025 export, the2026_1run the export byEDDYPRO_BIOMET_CH-LAE_2004-2026.09. EddyPro copies them into its output; the merge removes these copies. The meteorological columns of the dataset come from the meteo products. - Calibration gas. The CO2 concentrations of 14 December 2017 to 15 March 2019 were corrected before the runs (see Raw Data).
Merging the runs
Runs that meet at a year boundary overlap by 12 half-hours (31 December 19:15 to 1 January 00:45), and their values differ there, by up to 31.6 W m-2 in LE at the 2019/2020 boundary. Each record is taken whole from one run: before midnight from the earlier run, from midnight on from the later one. Columns without any value in the merged record are removed.